Class I Medical Devices: Requirements and CE Marking
What must manufacturers of class I medical devices comply with under the MDR? QMS, technical documentation, clinical evaluation, UDI, EUDAMED and PMS, step by step following MDCG 2019-15.
Nermin Duratović · Published · 6 min read
Class I is the lowest risk class for medical devices. In most cases, manufacturers may declare conformity themselves, without a notified body. This often creates the impression that class I involves little effort. It does not: the Medical Device Regulation (MDR) still requires a quality management system, complete technical documentation, a clinical evaluation and ongoing post-market surveillance. This article explains which obligations apply and in which order they can be implemented.
What is a class I medical device?
Classification follows the 22 rules in Annex VIII of the MDR. Class I mainly covers non-invasive devices and devices with short, low-criticality use, such as walking aids, wheelchairs or dressings for intact skin. Our article on MDR classification explains the process in detail.
Important: some products that were class I under the old directive (MDD) are classified higher under the MDR, for example many software products under Rule 11 or substance-based devices under Rule 21. An old classification should therefore always be reviewed.
Special cases: Is, Im and Ir
For three sub-groups, a notified body is involved, but only for a clearly limited aspect (Article 52(7) MDR):
| Sub-group | What the notified body reviews |
|---|---|
| Is – placed on the market sterile | Establishing, securing and maintaining sterile conditions |
| Im – with a measuring function | Conformity with metrological requirements |
| Ir – reusable surgical instruments | Reprocessing: cleaning, disinfection, sterilisation, maintenance, functional testing and instructions for use |
For this, the manufacturer applies either Chapters I and III of Annex IX (QMS assessment) or Part A of Annex XI (production quality assurance).
Obligations at a glance
Guidance document MDCG 2019-15 rev. 1 sets out the route for class I manufacturers in clear steps. In summary:
| Obligation | Applies to class I? | Legal basis |
|---|---|---|
| Quality management system | yes | Art. 10(9) |
| Risk management system | yes | Art. 10(2), Annex I |
| Clinical evaluation | yes | Art. 10(3), Art. 61, Annex XIV |
| Technical documentation | yes | Art. 10(4), Annexes II and III |
| EU declaration of conformity and CE marking | yes | Art. 19, Art. 20 |
| UDI system | yes | Art. 27 |
| Registration in EUDAMED | yes | Art. 29, Art. 31 |
| Person responsible for regulatory compliance (PRRC) | yes | Art. 15 |
| Post-market surveillance report (PMS report) | yes | Art. 85 |
| Periodic safety update report (PSUR) | no, only from class IIa | Art. 86 |
| Vigilance: reporting serious incidents | yes | Art. 87 |
Step by step to CE marking
1. Confirm medical device status and class
First, clarify whether the product is a medical device at all under Article 2(1) MDR. The intended purpose defined by the manufacturer is decisive. The class is then determined according to Annex VIII and the rationale documented.
2. Set up a QMS and risk management
Class I manufacturers also need a QMS that covers the requirements of Article 10(9). ISO 13485 certification is not mandatory for class I, but the standard is the usual framework. Risk management is generally carried out according to ISO 14971 and accompanies the device throughout its life cycle.
3. Meet the general safety and performance requirements
For each requirement in Annex I, record whether it applies, how it is met and where the evidence is. A GSPR checklist referencing standards, test reports and the risk analysis has proven effective.
4. Carry out a clinical evaluation
Clinical evaluation is mandatory for all risk classes. For established class I devices, it is often based on literature and experience data, but it must be prepared methodically in line with Annex XIV and updated regularly.
5. Prepare the technical documentation
The technical documentation under Annexes II and III includes, among other things, the device description, intended purpose, labelling and instructions for use, design and manufacturing information, risk management, verification and validation, and the PMS plan. Instructions for use and labelling must be available in the languages of the respective target markets.
6. Ensure general manufacturer obligations
These include in particular:
- PRRC: a person responsible for regulatory compliance under Article 15. Micro and small enterprises do not have to employ this person themselves, but must have them permanently and continuously at their disposal.
- Liability coverage for potential damage caused by defective devices (Art. 10(16)).
- Authorised representative in the EU if the manufacturer is based outside the Union.
7. Declaration of conformity and CE marking
Only once all evidence is in place does the manufacturer draw up the EU declaration of conformity according to Annex IV and affix the CE marking. For plain class I devices without Is, Im or Ir, no notified body number appears next to the CE mark.
8. UDI and EUDAMED
The manufacturer registers in EUDAMED and receives a single registration number (SRN). Each device is assigned a Basic UDI-DI and recorded in the database. The UDI enables traceability down to the individual device or batch. Our article on EUDAMED registration, obligations and deadlines explains which deadlines apply and how registration works.
9. Post-market surveillance
Surveillance starts when the device is placed on the market: evaluate experience and complaints, update the PMS report under Article 85 when necessary and provide it to the authority on request, report serious incidents and field safety corrective actions, and control nonconforming products. The results feed back into risk management and clinical evaluation.
Common mistakes with class I devices
- Class carried over instead of reviewed: many former class I devices are classified higher under the MDR.
- Clinical evaluation treated as a formality: a brief reference to "proven technology" is not enough.
- Claims beyond the intended purpose: Article 7 MDR prohibits claims that attribute properties the device does not have or suggest other uses.
- Forgetting PMS after launch: the PMS report is mandatory for class I too and must be kept up to date.
- No PRRC appointed: small companies in particular often overlook this obligation.
Conclusion
Class I does not mean "no rules", it means "no notified body". The manufacturer bears sole responsibility for conformity and must be able to demonstrate it at any time if the authority checks. Building the QMS, technical documentation, clinical evaluation and PMS systematically from the start prepares the company well, including for the day the portfolio grows into higher risk classes and ISO 13485 certification becomes relevant.
We support manufacturers with classification, setting up the QMS and technical documentation, and preparing for CE marking. Learn more about our MedTech consulting.