Biocompatibility under ISO 10993: A Guide for Manufacturers

How the biological evaluation of medical devices under ISO 10993 works: body contact, contact duration, endpoints, chemical characterisation and what is new in ISO 10993-1:2025.

Nermin Duratović · Published · 5 min read

Every medical device that comes into contact with the human body must be biologically safe. Whether catheter, implant, pen housing or wound dressing: the material must not cause unacceptable reactions such as irritation, allergies or toxic effects. How this is demonstrated is governed by the ISO 10993 series of standards. For manufacturers as well as suppliers of plastic and metal components, it is one of the most important sets of rules. This article explains how the biological evaluation works and what changes with the new edition of ISO 10993-1.

What "biocompatibility" means

Biocompatibility is not a material property that is tested once and ticked off. It depends on the finished device and on how and for how long it is in contact with the body. The same polymer may be uncritical for a briefly used skin-contact product and require extensive evidence for a long-term implant. Processing, colourants, release agents, cleaning and sterilisation also affect the result.

The Medical Device Regulation (MDR) requires in Annex I that risks posed by substances that may leach from the device are reduced as far as possible. ISO 10993 is the recognised way to demonstrate this.

Biological evaluation as part of risk management

The central standard is ISO 10993-1. It requires the biological evaluation to be carried out within the risk management process according to ISO 14971. At the end of 2025, the sixth edition, ISO 10993-1:2025, was published. It anchors this approach even more firmly:

  • Biological hazards, hazardous situations and harms are explicitly considered following the logic of ISO 14971.
  • There are dedicated clauses on biological risk estimation, risk evaluation and risk control, and on the biological evaluation report.
  • The evaluation must be kept up to date throughout the life cycle, including with production and post-market data.
  • The standard provides new guidance on determining exposure duration and on materials characterisation.

Which edition your notified body expects, and which European version is harmonised under the MDR, should be clarified early in the project.

Step 1: Categorise the device

The evaluation starts by categorising the nature and duration of body contact:

Nature of contact Examples
Surface contact intact skin, mucosal membrane, breached surfaces
Externally communicating contact with blood path, tissue, bone, dentin, e.g. via infusion sets or endoscopes
Implant tissue, bone, blood
Contact duration Classic categories
limited up to 24 hours
prolonged more than 24 hours up to 30 days
long-term more than 30 days

For repeated use, the cumulative duration counts. The 2025 edition clarifies how exposure duration should be determined in such cases.

Step 2: Evaluate existing data

Before testing, existing information is gathered: material data, experience with identical materials and processes, literature and previous tests. The standard explicitly requires unnecessary animal testing to be avoided. Endpoints can often be justified with existing data or chemical characterisation.

Step 3: Chemical characterisation

Chemical characterisation according to ISO 10993-18 is usually the starting point today. Extraction studies determine which substances may be released from the device (extractables). A toxicological risk assessment according to ISO 10993-17 then checks whether the amounts found are safe for patients. This often makes it possible to replace biological tests or plan them more precisely.

Step 4: Biological endpoints

Depending on the nature and duration of contact, different biological effects (endpoints) must be assessed. The most important have their own parts of the series:

Endpoint Standard
Cytotoxicity ISO 10993-5
Sensitisation ISO 10993-10
Irritation ISO 10993-23
Systemic toxicity ISO 10993-11
Genotoxicity ISO 10993-3
Implantation effects ISO 10993-6
Interactions with blood ISO 10993-4
Sample preparation ISO 10993-12

Tests are performed on samples representative of the finished device, i.e. made from production material using the production process, cleaned and sterilised as in use.

Step 5: Biological evaluation report

All results, justifications and residual risks are brought together in the biological evaluation report. It is part of the technical documentation and is updated when changes occur.

What suppliers need to know

For injection moulders, extruders and component manufacturers, ISO 10993 is particularly relevant, because any change can affect the customer's biological evaluation:

  • different resin or resin manufacturer
  • new colourant or additive
  • changed process parameters, such as higher temperatures
  • new release agent or cleaning chemistry
  • different sterilisation method

That is why quality agreements follow the rule: notify first, then change (more in Becoming a MedTech supplier). Suppliers who can provide material data, raw material manufacturers' declarations and complete traceability give their customers a real advantage. For parts in contact with medicinal products, questions about extractables and leachables are added (see Injection moulding for medical devices and drug delivery).

Common mistakes

  • Only the raw material considered: a material certificate does not replace an evaluation of the finished device.
  • Testing without a plan: tests are commissioned "just in case" instead of being derived from risk assessment and chemical characterisation.
  • Changes not assessed: material or process changes are implemented without reviewing the biological evaluation.
  • Evaluation not updated: new editions of standards, complaints or literature are not taken into account.

Conclusion

Biocompatibility is not a one-off test but a risk-based evaluation of the finished device throughout its life cycle. With ISO 10993-1:2025, the link to risk management under ISO 14971 moves even further into focus. Combining chemical characterisation, existing data and targeted testing wisely saves time, cost and unnecessary animal testing.

We support manufacturers and suppliers in planning biological evaluations, handling material changes and setting up the necessary supplier processes. Learn more about our MedTech consulting.

Planning your biological evaluation?

We help with the evaluation strategy under ISO 10993 and clarify which tests are really needed.

Discuss your project

30 min · free · no obligation

More articles

  • MDR,
  • Class I,
  • CE Marking

Class I Medical Devices: Requirements and CE Marking

What must manufacturers of class I medical devices comply with under the MDR? QMS, technical documentation, clinical evaluation, UDI, EUDAMED and PMS, step by step following MDCG 2019-15.

· 6 min readRead more
  • EUDAMED,
  • MDR,
  • UDI

EUDAMED: Registration, Obligations and Deadlines – What Applies Now

EUDAMED has been mandatory since 28 May 2026, and devices already on the market must be registered by 28 November 2026 at the latest. What applies now, which modules are still to come and how SRN, UDI and Basic UDI-DI fit together.

· 6 min readRead more