Injection Moulding for Medical Devices and Drug Delivery

What plastic injection moulders and assembly companies must meet for medical devices and drug delivery systems: materials, validation, cleanroom, traceability and MDR Article 117.

Nermin Duratović · Published · 5 min read

Plastic components are found in almost every medical device: housings, needle hubs, valves, catheter parts and, above all, drug delivery systems such as pens, autoinjectors and inhalers. For injection moulders, toolmakers and assembly companies, this is a growing market with high volumes and long run times. At the same time, requirements are particularly strict, because parts may come into contact with patients or with the medicinal product. This article shows what matters in injection moulding and assembly for medical technology and drug delivery.

Who must meet which rules

The injection moulder is usually not a manufacturer within the meaning of the Medical Device Regulation (MDR). The obligations reach it through its customer: under ISO 13485, the manufacturer must select, evaluate and monitor its suppliers on a risk basis and remains responsible for the overall product. Our article Becoming a MedTech supplier explains what this means for suppliers in general.

Drug delivery systems have an additional feature: medical device law and pharmaceutical law meet.

Product type Example Regulation
Stand-alone medical device reusable pen, syringe without medicinal product MDR, CE marking by the manufacturer
Co-packaged applicator supplied in the medicine pack Pharmaceutical law for the medicine; the device needs its own CE marking
Integral combination prefilled pen or syringe, usable only together, not reusable Pharmaceutical law; under MDR Article 117, the marketing authorisation dossier must show that the device part meets the safety and performance requirements of MDR Annex I

For integral combination products, Article 117 requires the marketing authorisation application to include either a declaration of conformity or certificate for the device part or, if this is missing and the device part on its own would require a notified body, a notified body opinion. For the moulder, this means its processes and evidence indirectly end up in a medicinal product's marketing authorisation dossier. Pharmaceutical customers audit their suppliers accordingly.

For needle-based injection systems, the ISO 11608 series is also relevant, setting out requirements and test methods for pens, autoinjectors and their components.

Materials: selection, release, changes

Material selection in medical injection moulding is not purely a technical question:

  • Biocompatibility: parts in contact with the body must be biologically evaluated according to ISO 10993-1. The evaluation is part of risk management and relates to the finished part, i.e. material, processing, colourants and cleaning.
  • Contact with the medicinal product: for parts touching the drug, pharmaceutical customers expect data on extractables and leachables, i.e. substances that may migrate from the plastic into the medicine.
  • Sterilisability: not every polymer tolerates every sterilisation method. Ethylene oxide, gamma irradiation and electron beam affect polymers differently, for example their strength, colour or ageing.
  • Supply chain: resin producers change formulations or discontinue grades. Medical device customers expect such changes to be notified early, which is why many contracts secure supply and change notification from the resin supplier.

Ground rule: material, supplier, colour batch or additive are never changed without agreement from the customer.

Process validation: the core

Injection moulding is a classic special process: whether a part has internal stresses, voids or dimensional deviations cannot be fully checked on every part. Under ISO 13485 clause 7.5.6, the process must therefore be validated. A three-stage approach is common:

Stage Content in injection moulding
IQ (Installation Qualification) Machine, mould, temperature control units, dryers and peripherals correctly installed, calibrated and documented
OQ (Operational Qualification) Determine the process window, often using design of experiments (DoE); test parts at the limits of the window
PQ (Performance Qualification) Demonstrate that the process consistently produces good parts over several lots under production conditions

This includes defined process parameters, monitoring them during production, and a rule for when revalidation is required after changes to the mould, machine or material. More on validation and supplier obligations in Becoming a MedTech supplier.

Tooling and traceability

For multi-cavity moulds, customers often expect cavity-specific traceability, for example through a cavity number moulded into the part. This makes it possible to narrow down which cavity is affected if a defect occurs. It also includes:

  • batch traceability from resin to delivered part
  • documented mould maintenance and changes
  • first article inspection after every relevant mould change
  • record retention for at least as long as the customer requires for the lifetime of its device

Cleanroom and cleanliness

For parts with low particle and microbial load, especially if they are later packaged sterile, many moulders produce in a cleanroom according to ISO 14644-1, often ISO class 7 or 8. What matters is not only the room class but the overall concept:

  • particle and microbial monitoring with defined alert and action limits
  • personnel, material and waste flows
  • gowning and hygiene rules, training
  • machines and automation suitable for cleanroom use
  • packaging directly in the cleanroom so cleanliness is maintained up to the customer

Assembly and automation

For drug delivery systems, suppliers increasingly take on not just single parts but sub-assemblies and assembly. Automated assembly lines deliver reproducible quality but bring additional obligations:

  • validation of assembly processes such as joining, bonding, welding or spring assembly
  • validation of inspection systems, for example camera inspection or force-displacement monitoring. An automated inspection system must be shown to reliably detect defective parts.
  • validation of software that controls equipment or stores test results
  • functional testing according to customer requirements, for injection systems often based on ISO 11608

Checklist for getting started

  1. Define target products: which parts, which customers, with or without drug contact?
  2. Build a QMS according to ISO 13485 or extend the existing QMS (see ISO 13485 certification).
  3. Material strategy with medical-grade resins, supplier agreements and change notification.
  4. Validation concept for moulding, assembly, inspection systems and software.
  5. Assess cleanroom needs and qualify the cleanroom where required.
  6. Ensure traceability down to cavity and batch.
  7. Become audit-ready: customers and their notified bodies audit critical suppliers, including unannounced.

Conclusion

Injection moulding for medical devices and drug delivery is an attractive but demanding market. Companies that master material release, process validation, cleanroom operation and traceability become long-term partners for device and pharmaceutical manufacturers, often for the entire lifetime of a product.

From building an ISO 13485 certified MedTech service provider with an ISO class 7 cleanroom, we know these requirements first-hand. We support industrial companies from the first concept to the customer audit. Learn more about our business development and MedTech consulting.

Injection moulding for medical devices?

We support plastics processors in setting up ISO 13485, validation and traceability for MedTech customers.

Discuss your project

30 min · free · no obligation

More articles

  • Biocompatibility,
  • ISO 10993,
  • Industry

Biocompatibility under ISO 10993: A Guide for Manufacturers

How the biological evaluation of medical devices under ISO 10993 works: body contact, contact duration, endpoints, chemical characterisation and what is new in ISO 10993-1:2025.

· 5 min readRead more
  • MDR,
  • Class I,
  • CE Marking

Class I Medical Devices: Requirements and CE Marking

What must manufacturers of class I medical devices comply with under the MDR? QMS, technical documentation, clinical evaluation, UDI, EUDAMED and PMS, step by step following MDCG 2019-15.

· 6 min readRead more