All articles
  • ISO 14971,
  • Risk Management,
  • MDR

ISO 14971: Risk Management for Medical Devices

How risk management under ISO 14971 works: plan, risk analysis, evaluation, risk control, residual risk and post-production monitoring, plus what the MDR additionally requires.

Nermin Duratović · Published · 5 min read

Few standards are as central for medical device manufacturers as ISO 14971. It describes how the risks of a medical device are identified, evaluated, controlled and monitored throughout its life cycle. Almost every other process relies on it: design, usability, software, biocompatibility, clinical evaluation and post-market surveillance. This article explains the process step by step and shows where the Medical Device Regulation (MDR) goes beyond the standard.

Which edition applies?

The current edition is ISO 14971:2019, the third edition. In Europe it applies as EN ISO 14971:2019 + A11:2021 and is harmonised under the MDR. Amendment A11 adds a European annex (Annex ZA) showing how the standard relates to the general safety and performance requirements of the MDR.

In addition, there is the technical report ISO/TR 24971:2020. It contains no additional requirements but provides guidance and examples on applying ISO 14971, for instance on risk acceptability criteria or on evaluating overall residual risk.

Key terms

The standard works with a clear chain of terms:

Term Meaning Example
Hazard potential source of harm electrical energy, sharp edge, incorrect readings
Hazardous situation circumstance in which people are exposed to a hazard patient touches a defective, live housing
Harm injury or damage to health electric shock
Risk combination of probability and severity of harm "rare but serious"

The most common mistake in risk analyses is mixing these levels, for example entering "electric shock" as a hazard. Clean separation makes analyses traceable and audit-proof.

The process step by step

1. Risk management plan

A plan is prepared for each device. Among other things, it defines:

  • scope and life-cycle phases
  • responsibilities and authorities
  • risk acceptability criteria, derived from the company's risk policy
  • the method for evaluating overall residual risk
  • verification activities
  • activities for collecting production and post-production information

2. Risk analysis

The starting point is the intended use and reasonably foreseeable misuse. From these, safety-related characteristics, hazards and hazardous situations are derived, and the risk is estimated for each situation. Proven methods include preliminary hazard analysis (PHA), failure mode and effects analysis (FMEA) and fault tree analysis (FTA). The standard does not prescribe a specific method.

3. Risk evaluation

Each estimated risk is compared with the acceptability criteria in the plan. If a risk is not acceptable, risk control measures are required.

4. Risk control

Measures are selected in a fixed order of priority:

  1. Inherently safe design and manufacture, for example eliminating the hazard by design
  2. Protective measures in the device or manufacturing process, such as alarms or safeguards
  3. Information for safety and, where appropriate, user training

Each measure is implemented and its effectiveness verified. It must also be checked whether a measure introduces new hazards.

5. Overall residual risk

After all measures, the overall residual risk is evaluated, i.e. all remaining individual risks together. If it is not acceptable under the criteria in the plan, the manufacturer may assess whether the medical benefit outweighs the residual risk. Significant residual risks are disclosed in the accompanying information.

6. Risk management review

Before release for commercial distribution, the manufacturer reviews whether the plan has been implemented, the overall residual risk is acceptable and methods exist for collecting post-production information. The result is documented in the risk management report.

7. Production and post-production

Risk management does not end at launch. Information from production, complaints, incidents, literature and similar devices is collected and evaluated. If it reveals new hazards or higher risks, the risk management file is updated. This is where risk management and post-market surveillance under the MDR interlock.

What the MDR additionally requires

In Annex I, the MDR requires risks to be reduced as far as possible without adversely affecting the benefit-risk ratio. This is stricter than some in-house practices of reducing risks only "as low as reasonably practicable". Economic considerations must not play a role. The MDR also requires that

  • risk management runs as a continuous iterative process throughout the entire life cycle,
  • the results are linked with the clinical evaluation and post-market surveillance,
  • all residual risks are evaluated and the significant ones communicated to users.

The European Annex ZA of EN ISO 14971:2019 + A11:2021 shows where the standard covers MDR requirements and where additional considerations are needed.

Common audit findings

  • Acceptability criteria missing or not justified
  • Hazard, hazardous situation and harm mixed up
  • Risk controls not verified: a warning in the instructions for use is not a verified measure until its effectiveness is demonstrated.
  • Information for safety as first choice: the order of priority was skipped.
  • Risk management file not maintained after launch: complaints and incidents do not feed back.
  • Missing links: risk analysis, usability (IEC 62366-1), software (IEC 62304) and biocompatibility (ISO 10993) run side by side instead of together.

Conclusion

ISO 14971 is the backbone of every technical documentation. Well-maintained risk management connects design, manufacturing and market surveillance and makes decisions traceable, for the notified body as well as for your own team. Treating it as a living process from the start saves a lot of time in audits (see preparing for a notified body audit).

We support the setup and revision of risk management processes and risk management files according to ISO 14971 and the MDR. Learn more about our MedTech consulting.

Want your risk management reviewed?

We review your risk management file against ISO 14971 and show where notified bodies typically ask questions.

Discuss your project

30 min · free · no obligation

More articles

  • Biocompatibility,
  • ISO 10993,
  • Industry

Biocompatibility under ISO 10993: A Guide for Manufacturers

How the biological evaluation of medical devices under ISO 10993 works: body contact, contact duration, endpoints, chemical characterisation and what is new in ISO 10993-1:2025.

· 5 min readRead more
  • MDR,
  • Class I,
  • CE Marking

Class I Medical Devices: Requirements and CE Marking

What must manufacturers of class I medical devices comply with under the MDR? QMS, technical documentation, clinical evaluation, UDI, EUDAMED and PMS, step by step following MDCG 2019-15.

· 6 min readRead more