Software as a Medical Device: Qualification, Class, IEC 62304
When is software a medical device? Qualification under MDCG 2019-11, classification under Rule 11 and development according to IEC 62304, a guide for software and IT teams.
Nermin Duratović · Published · 6 min read
Apps, cloud platforms, image analysis algorithms: more and more software performs tasks that used to be reserved for physicians or hardware devices. For software companies and IT service providers, this raises a question that determines timeline and budget: is our software a medical device? And if so, what requirements follow? This article answers both in three steps: qualification, classification and development.
Step 1: Is the software a medical device?
Under the MDR, software is a medical device if the manufacturer gives it a medical intended purpose, such as diagnosis, monitoring, prediction, prognosis, treatment or alleviation of disease (Article 2(1)). What matters is not the technology but what the manufacturer says about the software: in the instructions for use, the app store description, the website and marketing.
Guidance document MDCG 2019-11 rev. 1, published in June 2025, describes the assessment in detail. It uses a decision tree, simplified:
- Is it software within the meaning of the guidance?
- Does it drive or influence a hardware medical device, or is it an accessory? Then it is regulated as part of that device or as an accessory.
- Does it perform an action on data beyond storage, archival, communication, simple search or lossless compression?
- Is that action for the benefit of individual patients?
- Does it meet the definition of a medical device under Article 2(1) MDR?
If the answer to steps 3 to 5 is "yes" in each case, the product is medical device software (MDSW).
| Usually a medical device | Usually not a medical device |
|---|---|
| Analyses patient data to support diagnostic or therapeutic decisions | Stores, archives or transmits data without analysing it |
| Monitors physiological values and flags anomalies | Scheduling, billing, hospital administration |
| Calculates doses for individual patients | General fitness and lifestyle apps without medical claims |
| Analyses medical images for findings | Software that only aggregates population data for statistics or research |
The 2025 revision stresses the importance of a clearly worded intended purpose and explains in more detail how modular software is handled. If only one module of a platform has a medical purpose, the medical device boundary can be drawn around that module. The interfaces to the other modules must then be clearly defined.
Step 2: Which risk class?
For software, Rule 11 in Annex VIII of the MDR is the key rule:
- Software that provides information used to take diagnostic or therapeutic decisions is class IIa.
- If those decisions may cause a serious deterioration of health or a surgical intervention: class IIb.
- If they may cause death or an irreversible deterioration: class III.
- Software intended to monitor physiological processes: class IIa, or class IIb for vital parameters where variations could result in immediate danger.
- All other software: class I.
In practice, most medical device software therefore lands in class IIa or higher, which means a notified body is involved. Our article on MDR classification explains how this interacts with the other rules. If the software includes AI, the EU AI Act also applies from class IIa.
Step 3: Development according to IEC 62304
The MDR requires software to be developed according to the state of the art, taking into account the life cycle, risk management, verification and validation (Annex I, Section 17.2). The recognised state of the art is IEC 62304 "Medical device software – Software life cycle processes".
The standard does not prescribe a method. Agile development is explicitly possible. It does, however, require documented processes for:
- software development planning
- software requirements analysis
- architecture and detailed design
- implementation and verification of software units
- integration and integration testing
- software system testing
- release
- maintenance, configuration management and problem resolution
Safety classes A, B and C
In the current edition, the amount of documentation required depends on the software safety class. It is determined by the most severe harm a software failure could contribute to:
| Safety class | Possible harm |
|---|---|
| A | no injury or damage to health possible |
| B | no serious injury possible |
| C | death or serious injury possible |
The higher the class, the more activities the standard requires, such as a detailed architecture, detailed design and verification at software unit level.
Important: the IEC 62304 safety class is not the same as the MDR risk class. A class IIa device may contain software of safety class B or C, and vice versa.
Outlook: IEC 62304 Edition 2
The standard is currently undergoing a fundamental revision. The drafts for the second edition propose, among other things, extending the scope to health software in general and replacing the three safety classes with two "process rigour levels". Reports on the publication status differ, and even once the new edition appears, a transition period is to be expected. Teams developing today continue to work to the current edition and should set up their processes so they can adapt to the new logic later.
What else is needed beyond IEC 62304
Standards-compliant development alone is not enough for CE marking. Manufacturers of medical device software also need:
- a quality management system, usually according to ISO 13485,
- risk management according to ISO 14971, linked to the software risk analysis,
- usability engineering according to IEC 62366-1,
- cybersecurity to the state of the art, usually implemented according to IEC 81001-5-1 (see medical device cybersecurity),
- a clinical evaluation, including for software,
- technical documentation according to Annexes II and III MDR,
- post-market surveillance, including the assessment of updates.
Updates and changes
Software changes constantly, while medical devices should only change in a controlled way. Every change must therefore be assessed: does it affect the intended purpose, safety or performance? Substantial changes to devices of class IIa and above must be notified to the notified body before they are implemented. Well-designed change management with clear criteria prevents every update from becoming a regulatory project.
Checklist for software and IT teams
- Define the intended purpose in writing and align all texts with it, including marketing and app store listings.
- Document the qualification according to MDCG 2019-11, even if the result is "not a medical device".
- Determine the class under Rule 11 and justify it.
- Draw module boundaries deliberately if only parts of the platform are medical.
- Align the development process with IEC 62304 and determine the safety class.
- Record and assess SOUP and open-source components. IEC 62304 calls third-party software "software of unknown provenance" (SOUP).
- Talk to a notified body early if the software is class IIa or higher.
Conclusion
Whether software is a medical device depends on its intended purpose, not on its technology. Documenting qualification and classification early and carefully avoids costly surprises. For development itself, IEC 62304 provides a clear framework that combines well with agile methods.